Comparison of image quality, myocardial perfusion, and left ventricular function between standard imaging and single-injection ultra-low-dose imaging using a high-efficiency SPECT camera: the MILLISIEVERT study.
Level II
Where this comes from
- Record sourced from PubMed, PMID 24982439.
- Also identified by DOI 10.2967/jnumed.114.138222 and PMC identifier 4486330.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
SPECT myocardial perfusion imaging plays a central role in coronary artery disease diagnosis, but concerns exist regarding its radiation burden. Compared with standard Anger SPECT (A-SPECT) cameras, new high-efficiency (HE) cameras with specialized collimators and solid-state cadmium-zinc-telluride detectors offer potential to maintain image quality (IQ), while reducing administered activity and thus radiation dose to patients. No previous study has compared IQ, interpretation, total perfusion deficit (TPD), or ejection fraction (EF) in patients receiving both ultra-low-dose (ULD) imaging on an HE SPECT camera and standard low-dose (SLD) A-SPECT imaging. We compared ULD HE SPECT with SLD A-SPECT imaging by dividing the rest dose in 101 patients at 3 sites scheduled to undergo clinical A-SPECT myocardial perfusion imaging using a same day rest-stress (99m)Tc protocol. Patients underwent HE SPECT imaging after an initial approximately 130-MBq (3.5 mCi) dose and SLD-A-SPECT imaging after the remainder of the planned dose. Images were scored visually by 2 masked readers for IQ and summed rest score. TPD and EF were assessed quantitatively. Mean activity was 134 MBq (3.62 mCi) for ULD HE SPECT (effective dose, 1.15 mSv) and 278 MBq (7.50 mCi, 2.39 mSv) for SLD A-SPECT. Overall IQ was superior for ULD HE SPECT (P < 0.0001), with twice as many studies graded excellent quality. Extracardiac activity and overall perfusion assessment were similar. Between-method correlations were high for summed rest score (r = 0.87), TPD (r = 0.91), and EF (r = 0.88). ULD HE SPECT rest imaging correlates highly with SLD A-SPECT. It has improved image quality, comparable extracardiac activity, and achieves radiation dose reduction to 1 mSv for a single injection.
Medical subject headings
- Coronary Artery Disease
- Myocardial Perfusion Imaging
- Tomography, Emission-Computed, Single-Photon
- Ventricular Function, Left