Reproducible fixation with a tapered, fluted, modular, titanium stem in revision hip arthroplasty at 8-15 years follow-up.

Rodriguez, Jose A; Deshmukh, Ajit J; Robinson, Jonathan; Cornell, Charles N; Rasquinha, Vijay J; Ranawat, Amar S; Ranawat, Chitranjan S · J Arthroplasty · 2014

retrospective_cohort · Level III

Where this comes from

Abstract

The use of tapered, fluted, modular, distally fixing stems has increased in femoral revision surgery. The goal of this retrospective study was to assess mid-term to long-term outcomes of this implant. Seventy-one hips in 70 patients with a mean age of 69 years were followed for an average of 10 years. Preoperative HHS averaged 50 and improved to 87 postoperatively. Seventy-nine percent hips had Paprosky type 3A or more bone-loss. All stems osseointegrated distally (100%). Two hips subsided >5mm but achieved secondary stability. Sixty-eight percent hips had evidence of bony reconstitution and 21% demonstrated diaphyseal stress-shielding. One stem fractured near its modular junction and was revised with a mechanical failure rate of 1.4%. Distal fixation and clinical improvement were reproducibly achieved with this stem design.

Medical subject headings

Anatomy