Quality control within the multicentre perfusion CT study of primary colorectal cancer (PROSPeCT): results of an iodine density phantom study.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 25001085.
- Also identified by DOI 10.1007/s00330-014-3258-y.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To assess the cross-centre consistency of iodine enhancement, contrast-to-noise ratio and radiation dose in a multicentre perfusion CT trial of colorectal cancer. A cylindrical water phantom containing different iodine inserts was examined on seven CT models in 13 hospitals. The relationship between CT number (Hounsfield units, HU) and iodine concentration (milligrams per millilitre) was established and contrast-to-noise ratios (CNRs) calculated. Radiation doses (CTDIvol, DLP) were compared across all sites. There was a linear relationship between CT number and iodine density. Iodine enhancement varied by a factor of at most 1.10, and image noise by at most 1.5 across the study sites. At an iodine concentration of 1 mg ml(-1) and 100 kV, CNRs ranged from 3.6 to 4.8 in the 220-mm phantom and from 1.4 to 1.9 in the 300-mm phantom. Doses varied by a factor of at most 2.4, but remained within study dose constraints. Iterative reconstruction algorithms did not alter iodine enhancement but resulted in reduced image noise by a factor of at most 2.2, allowing a potential dose decrease of at most 80% compared to filtered back projection (FBP). Quality control of CT performance across centres indicates that CNR values remain relatively consistent across all sites, giving acceptable image quality within the agreed dose constraints. Quality control is essential in a multicentre setting to enable CT quantification. CNRs in a body-sized phantom had the recommended value of at least 1.5. CTDIs and DLPs varied by factors of 1.8 and 2.4 respectively.
Medical subject headings
- Algorithms
- Colorectal Neoplasms
- Iodine
- Phantoms, Imaging
- Quality Control
- Radiographic Image Interpretation, Computer-Assisted
- Tomography, X-Ray Computed