Tibial rotational alignment was significantly improved by use of a CT-navigated control device in total knee arthroplasty.
Where this comes from
- Record sourced from PubMed, PMID 25070900.
- Also identified by DOI 10.1016/j.arth.2014.06.016.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study compared the accuracy of three methods to set tibial component rotational alignment: (1) conventional method, the anteroposterior (AP) axis was determined by the surgeon using anatomical landmarks; (2) partial-navigation method, the tibia was prepared according to the AP axis using a CT-based navigation system and the component was manually positioned; (3) full-navigation method, the tibial component was positioned and fixed with cement under the control of navigation using a newly developed instrument. The conventional method showed considerable deviation (range, -18.6° to 14.7°), and the partial-navigation method also showed considerable deviation (-11.3° to 8.1°). In contrast, the full-navigation method significantly improved the accuracy of alignment (-2.9° to 2.1°). The tibial component can become malaligned during cement fixation, even after proper bone preparation.