Remote detection of photoplethysmographic systolic and diastolic peaks using a digital camera.
Where this comes from
- Record sourced from PubMed, PMID 25073159.
- Also identified by DOI 10.1109/TBME.2014.2340991.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We present a new method for measuring photoplethysmogram signals remotely using ambient light and a digital camera that allows for accurate recovery of the waveform morphology (from a distance of 3 m). In particular, we show that the peak-to-peak time between the systolic peak and diastolic peak/inflection can be automatically recovered using the second-order derivative of the remotely measured waveform. We compare measurements from the face with those captured using a contact fingertip sensor and show high agreement in peak and interval timings. Furthermore, we show that results can be significantly improved using orange, green, and cyan color channels compared to the tradition red, green, and blue channel combination. The absolute error in interbeat intervals was 26 ms and the absolute error in mean systolic-diastolic peak-to-peak times was 12 ms. The mean systolic-diastolic peak-to-peak times measured using the contact sensor and the camera were highly correlated, ρ = 0.94 (p 0.001). The results were obtained with a camera frame-rate of only 30 Hz. This technology has significant potential for advancing healthcare.
Medical subject headings
- Blood Pressure Determination
- Blood Volume Determination
- Face
- Photography
- Photoplethysmography