Cross polarization compatible dialysis chip.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 25105977.
- Also identified by DOI 10.1039/c4lc00600c.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We visualize birefringence in microliter sample volumes using a microfluidic dialysis chip optimized for cross polarization microscopy. The chip is composed of two overlapping polydimethylsiloxane (PDMS) channels separated by a commercial cellulose ester membrane. Buffer exchange in the sample chamber is achieved within minutes by dialyzing under continuous reservoir flow. Using fd virus as a birefringent model system, we monitor the fd virus isotropic to liquid crystal phase transition as a function of ionic strength. We show that the reorientation of the fd virus spans a few tens of seconds, indicative of fast ion exchange across the membrane. Complete phase separation reorganization takes minutes to hours as it involves diffusive virus mass transport within the storage chamber.
Medical subject headings
- Microdialysis
- Microfluidic Analytical Techniques