A biomechanical comparison of internal fixation techniques for ankle arthrodesis.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 25116232.
- Also identified by DOI 10.1053/j.jfas.2014.06.002.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of the present study was to compare the primary bending stiffness characteristics of 5 different ankle arthrodesis fixation techniques: 3 compression screws, an anterior locking plate, a lateral locking plate, an anterior locking plate with a compression screw, and a lateral locking plate with a compression screw. A total of 25 full-scale anatomic models consisting of fourth-generation composite tibiae and tali were tested using an Instron 4505 Universal Testing System. We hypothesized that the use of a compression screw with a locking plate would add considerable stiffness to the fixation construct compared with the use of a locking plate alone. The data have shown that an anterior or lateral plate with a compression screw provides significantly greater stiffness than both a plate and 3 compression screws used individually. No significant difference was seen between the anterior plate with a compression screw and the lateral plate with a compression screw. No significant differences were found among the use of an anterior plate, a lateral plate, or 3 compression screws. We have concluded that when using a locking plate in an anterior or lateral configuration, the addition of a compression screw will considerably increase the primary bending stiffness of ankle arthrodesis.
Medical subject headings
- Ankle Joint
- Arthrodesis
- Internal Fixators
Anatomy
- ankle