Dual-energy liver CT: effect of monochromatic imaging on lesion detection, conspicuity, and contrast-to-noise ratio of hypervascular lesions on late arterial phase.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 25148163.
- Also identified by DOI 10.2214/AJR.13.11337.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to evaluate the effect of use of dual-energy CT monochromatic imaging in the late hepatic arterial phase on hyperenhancing focal lesion detection and lesion conspicuity. This prospective study included 72 patients imaged with a single-source dual-energy CT scanner. Late arterial phase imaging was performed with dual energies of 140 and 80 kVp, and the portal venous and delayed phases were performed with a single energy of 120 kVp. Two deidentified image sets were created: set A consisted of 77-keV images only, and set B consisted of 40-, 50-, 70-, and 77-keV images and iodine-based contrast material decomposition images. Two independent reviewers identified hypervascular lesions and subjectively scored lesion conspicuity. Contrast-to-noise ratios were calculated, and radiation dose (volume CT dose index) was recorded. The 128 lesions identified had a mean size of 1.7 ± 1.4 cm. There was no difference in lesion detection between the two reviewers or the two image sets. The contrast-to-noise ratio at 50 keV was 72% greater than that at 77 keV (p < 0.0001). Subjective conspicuity was statistically greatest at 50 keV (p < 0.0001). There was no statistical difference in mean volume CT dose index between the dual-energy (12.8 mGy) and the two single-energy (14.4 and 14.2 mGy) phases. Viewing dual-energy CT images may result in the greatest subjective lesion conspicuity and measured contrast-to-noise ratio at 50 keV with equal detection of hyperenhancing liver lesions compared with viewing 77-keV images alone. In addition, the radiation doses of dual-energy CT may be similar to those of single-energy CT.
Medical subject headings
- Liver Cirrhosis
- Liver Neoplasms
- Neovascularization, Pathologic
- Precancerous Conditions
- Radiographic Image Interpretation, Computer-Assisted
- Radiography, Dual-Energy Scanned Projection