A novel method for implementation of frameless StereoEEG in epilepsy surgery.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 25161004.
- Also identified by DOI 10.1227/NEU.0000000000000544 and PMC identifier 4224568.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Stereoelectroencephalography (SEEG) is an invasive diagnostic procedure in epilepsy surgery that is usually implemented with frame-based methods. To describe a new technique of frameless SEEG and report a prospective case series at a single center. Image integration and planning of electrode trajectories were performed preoperatively on specialized software and exported to a Medtronic S7 StealthStation. Trajectories were implemented by frameless stereotaxy using percutaneous drilling and bolt insertion. Twenty-two patients went this technique, with the insertion of 187 intracerebral electrodes. Of 187 electrodes, 175 accurately reached their neurophysiological target, as measured by postoperative computed tomography reconstruction and multimodal image integration with preoperative magnetic resonance imaging. Four electrodes failed to hit their target due to extradural deflection, and 3 were subsequently resited satisfactorily. Eight electrodes were off target by a mean of 3.6 mm (range, 0.9-6.8 mm) due to a combination of errors in bolt trajectory implementation and bending of the electrode. There was 1 postoperative hemorrhage that was clinically asymptomatic and no postoperative infections. Sixteen patients were offered definitive cortical resections, and 6 patients were excluded from resective surgery. Frameless SEEG is a novel and safe method for implementing SEEG and is easily translated into clinical practice.
Medical subject headings
- Electrodes, Implanted
- Electroencephalography
- Epilepsy
- Neuronavigation