Comparison of two pressurisers for cementation of the proximal femur.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 25163947.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To compare pressures generated by 2 different cement pressurisers at various locations in the proximal femur. Two groups of 5 synthetic femurs were used, and 6 pressure sensors were placed in the femur at 20-mm intervals proximally to distally. Cement was filled into the femoral canal retrogradely using a cement gun with either the half-moon pressuriser or the femoral canal pressuriser. Maximum pressures and pressure time integrals (cumulative pressure over time) of the 2 pressurisers were compared. At all sensors, the half-moon pressuriser produced higher maximum pressures and pressure time integrals than the femoral canal pressuriser, but the difference was significant only at sensor 1 (proximal femur). This may result in reduced cement interdigitation in the proximal femur. The half-moon pressuriser produced higher maximum cementation pressures and pressure time integrals than the femoral canal pressuriser in the proximal femur region, which is critical for rotational stability of the implant and prevention of implant fracture.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Cementation
- Femur
- Hip Prosthesis
- Pressure
Anatomy
- femur
- hip