Clinical results of the conserve plus metal on metal hip resurfacing: an independent series.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 25212284.
- Also identified by DOI 10.1016/j.arth.2014.08.005.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of the present study was to report the clinical and radiographic results of an independent series of the Conserve Plus hip resurfacing. Five hundred forty-eight consecutive hip resurfacings were performed using the Conserve Plus prosthesis in 458 patients (350 males) with a mean age of 48.3 years (range 19 to 66). No patients were lost to follow-up. At a mean follow-up of 6.6 years (3.9 to 11.9) thirty (5.4%) hips required conversion to a total hip arthroplasty (THA) (20 males, 10 females, mean age=48.3±7.3 years). Five-year survival with as revision endpoint was 94.5% (95% CI: 93.5% to 95.5%). This study confirms the good clinical results previously reported with the Conserve Plus hip resurfacing device.
Medical subject headings
- Arthritis
- Arthroplasty, Replacement, Hip
- Hip Joint
- Hip Prosthesis
Anatomy
- hip
- femur