Pyrocarbon interposition (PyroDisk) implant for trapeziometacarpal osteoarthritis: minimum 5-year follow-up.

Barrera-Ochoa, Sergi; Vidal-Tarrason, Nuria; Correa-Vázquez, Eva; Reverte-Vinaixa, Maria Mercè; Font-Segura, Jordi; Mir-Bullo, Xavier · J Hand Surg Am · 2014

retrospective_cohort · Level III

Where this comes from

Abstract

To report outcomes after a minimum of 5 years following pyrocarbon interposition (PyroDisk) trapeziometacarpal joint implant for osteoarthritis at a single center. We retrospectively reviewed the midterm clinical and radiological outcomes of 19 patients who had a pyrocarbon interposition implant (PyroDisk; Integra Life Sciences, Plainsboro, NJ) arthroplasty. The rate and causes of repeat surgeries, revisions, and complications were examined. The mean follow-up period was 68 months. Patient satisfaction was high. The mobility of the operated thumb was restored to a range of motion comparable with that of the contralateral thumb. Grip strength improved by 26%. Overall function, according to the Quick Disabilities of the Arm, Shoulder, and Hand score, showed an average improvement of 71 to 20. Pain decreased by 78% according to the numerical rating scale. Radiological evaluation using a modification of the system described by Herren revealed progression of the periprosthetic lucency (grade I-II) of the implant after 5 years in 5 of 19 (26%) patients. Progression of lucency did not predict implant loosening or failure at 5 years. Two patients had symptomatic instability that required revision. No dislocations occurred. The 5-year survival of the prosthesis was 90%. The PyroDisk implant for treating advanced trapeziometacarpal arthritis did not demonstrate superiority over published outcome data of trapeziectomy with or without ligament reconstruction and tendon interposition. Therapeutic IV.

Medical subject headings

Anatomy