Dixon imaging-based partial volume correction improves quantification of choline detected by breast 3D-MRSI.

Minarikova, Lenka; Gruber, Stephan; Bogner, Wolfgang; Pinker-Domenig, Katja; Baltzer, Pascal A T; Helbich, Thomas H; Trattnig, Siegfried; Chmelik, Marek · Eur Radiol · 2015

other · Level V

Where this comes from

Abstract

Our aim was to develop a partial volume (PV) correction method of choline (Cho) signals detected by breast 3D-magnetic resonance spectroscopic imaging (3D-MRSI), using information from water/fat-Dixon MRI. Following institutional review board approval, five breast cancer patients were measured at 3 T. 3D-MRSI (1 cm(3) resolution, duration ~11 min) and Dixon MRI (1 mm(3), ~2 min) were measured in vivo and in phantoms. Glandular/lesion tissue was segmented from water/fat-Dixon MRI and transformed to match the resolution of 3D-MRSI. The resulting PV values were used to correct Cho signals. Our method was validated on a two-compartment phantom (choline/water and oil). PV values were correlated with the spectroscopic water signal. Cho signal variability, caused by partial-water/fat content, was tested in 3D-MRSI voxels located in/near malignant lesions. Phantom measurements showed good correlation (r = 0.99) with quantified 3D-MRSI water signals, and better homogeneity after correction. The dependence of the quantified Cho signal on the water/fat voxel composition was significantly (p < 0.05) reduced using Dixon MRI-based PV correction, compared to the original uncorrected data (1.60-fold to 3.12-fold) in patients. The proposed method allows quantification of the Cho signal in glandular/lesion tissue independent of water/fat composition in breast 3D-MRSI. This can improve the reproducibility of breast 3D-MRSI, particularly important for therapy monitoring.

Medical subject headings