Prediction of inflammatory responses induced by biomaterials in contact with human blood using protein fingerprint from plasma.

Engberg, Anna E; Nilsson, Per H; Huang, Shan; Fromell, Karin; Hamad, Osama A; Mollnes, Tom Eirik; Rosengren-Holmberg, Jenny P; Sandholm, Kerstin et al. · Biomaterials · 2015

basic_science · Level V

Where this comes from

Abstract

Inappropriate complement activation is often responsible for incompatibility reactions that occur when biomaterials are used. Complement activation is therefore a criterion included in legislation regarding biomaterials testing. However, no consensus is yet available regarding appropriate complement-activation-related test parameters. We examined protein adsorption in plasma and complement activation/cytokine release in whole blood incubated with well-characterized polymers. Strong correlations were found between the ratio of C4 to its inhibitor C4BP and generation of 10 (mainly pro-inflammatory) cytokines, including IL-17, IFN-γ, and IL-6. The levels of complement activation products correlated weakly (C3a) or not at all (C5a, sC5b-9), confirming their poor predictive values. We have demonstrated a direct correlation between downstream biological effects and the proteins initially adhering to an artificial surface after contact with blood. Consequently, we propose the C4/C4BP ratio as a robust, predictor of biocompatibility with superior specificity and sensitivity over the current gold standard.

Medical subject headings