Retinal thickness measurement obtained with spectral domain optical coherence tomography assisted optical biopsy accurately correlates with ex vivo histology.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 25360629.
- Also identified by DOI 10.1371/journal.pone.0111203 and PMC identifier 4216007.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
BACKGROUND: This study determines 'correlation constants' between the gold standard histological measurement of retinal thickness and the newer spectral-domain optical coherence tomography (SD-OCT) technology in adult C57BL/6 mice. METHODS: Forty-eight eyes from adult mice underwent SD-OCT imaging and then were histologically prepared for frozen sectioning with H&E staining. Retinal thickness was measured via 10x light microscopy. SD-OCT images and histological sections were standardized to three anatomical sites relative to the optic nerve head (ONH) location. The ratios between SD-OCT to histological thickness for total retinal thickness (TRT) and six sublayers were defined as 'correlation constants'. RESULTS: Mean (± SE) TRT for SD-OCT and histological sections was 210.95 µm (± 1.09) and 219.58 µm (± 2.67), respectively. The mean 'correlation constant' for TRT between the SD-OCT and histological sections was 0.96. The retinal thickness for all sublayers measured by SD-OCT vs. histology were also similar, the 'correlation constant' values ranged from 0.70 to 1.17. All SD-OCT and histological measurements demonstrated highly significant (p<0.01) strong positive correlations. CONCLUSION: This study establishes conversion factors for the translation of ex vivo data into in vivo information; thus enhancing the applicability of SD-OCT in translational research.
Medical subject headings
- Biopsy
- Histological Techniques
- Retina
- Tomography, Optical Coherence