The effects of Patent Blue dye on peripheral and cerebral oxyhaemoglobin saturations.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 25389044.
- Also identified by DOI 10.1111/anae.12932.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We measured the effect of Patent Blue dye on oxyhaemoglobin saturations after injection into breast tissue: 40 women had anaesthesia for breast surgery maintained with sevoflurane or propofol (20 randomly allocated to each). Saturations were recorded with a digital pulse oximeter, in arterial blood samples and with a cerebral tissue oximeter before dye injection and 10, 20, 30, 40, 50, 60, 75, 90, 105 and 120 min afterwards. Patent Blue did not decrease arterial blood oxyhaemoglobin saturation, but it did reduce mean (SD) digital and cerebral oxyhaemoglobin saturations by 1.1 (1.1) % and 6.8 (7.0) %, p < 0.0001 for both. The falsely reduced oximeter readings persisted for at least 2 h. The mean (SD) intra-operative digital pulse oxyhaemoglobin readings were lower with sevoflurane than propofol, 97.8 (1.2) % and 98.8 (1.0) %, respectively, p < 0.0001.