To determine if the manner in which a tuning fork is activated affects its vibrational response. Diagnostic test assessment. Hearing Center of Excellence laboratory. A Polytec OFV-5000 scanning vibrometer was used to measure the vibrational response of 256-Hz, 512-Hz, and 1024-Hz tuning forks after activation. The tuning forks were activated to varying intensities by striking 4 unique surfaces: the head, palm, a metal surface, and a wood table. The fundamental frequency of the individual tuning fork was the dominant observed frequency in all testing scenarios. Additional nonharmonic frequencies were noted when the 256-Hz and 512-Hz tuning forks were struck off metal and wooden surfaces. Additional nonfundamental sound frequencies produced secondary to striking a tuning fork off a metal object or a wooden table could affect clinical tuning fork examination and complicate decisions regarding surgical candidacy.
Medical subject headings
- Bone Conduction
- Hearing
- Hearing Loss, Conductive
- Hearing Tests
- Otolaryngology