Intraoperative navigation of patient-specific instrumentation does not predict final implant position.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 25486944.
- Also identified by DOI 10.1016/j.arth.2014.11.005.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to determine if intraoperative navigation predicted final implant position of total knee arthroplasties (TKAs) performed with patient-specific instrumentation (PSI). We retrospectively reviewed 60 TKAs performed with PSI and imageless navigation. These values were compared to postoperative coronal alignment based on long-leg radiographs, as well as rotation and tibial slope based on CT scans. The intraoperative coronal position of the tibia as measured by the intraoperative navigation indicated a significantly higher deviation from the neutral mechanical axis than the actual final position (P=0.03). Similarly, tibial slope and femoral component rotation measured by intraoperative navigation significantly deviated from the final slope and femoral component rotation (P<0.0001). In conclusion, intraoperative navigation of PSI position showed a significantly high deviation from the true final implant position. Level III, therapeutic. See Instructions to Authors for a complete description of levels of evidence.