Patient-specific targeting guides compared with traditional instrumentation for glenoid component placement in shoulder arthroplasty: a multi-surgeon study in 70 arthritic cadaver specimens.

Throckmorton, Thomas W; Gulotta, Lawrence V; Bonnarens, Frank O; Wright, Stephen A; Hartzell, Jeffrey L; Rozzi, William B; Hurst, Jason M; Frostick, Simon P et al. · J Shoulder Elbow Surg · 2015

basic_science · Level V

Where this comes from

Abstract

The purpose of this study was to compare the accuracy of patient-specific guides for total shoulder arthroplasty (TSA) with traditional instrumentation in arthritic cadaver shoulders. We hypothesized that the patient-specific guides would place components more accurately than standard instrumentation. Seventy cadaver shoulders with radiographically confirmed arthritis were randomized in equal groups to 5 surgeons of varying experience levels who were not involved in development of the patient-specific guidance system. Specimens were then randomized to patient-specific guides based off of computed tomography scanning, standard instrumentation, and anatomic TSA or reverse TSA. Variances in version or inclination of more than 10° and more than 4 mm in starting point were considered indications of significant component malposition. TSA glenoid components placed with patient-specific guides averaged 5° of deviation from the intended position in version and 3° in inclination; those with standard instrumentation averaged 8° of deviation in version and 7° in inclination. These differences were significant for version (P = .04) and inclination (P = .01). Multivariate analysis of variance to compare the overall accuracy for the entire cohort (TSA and reverse TSA) revealed patient-specific guides to be significantly more accurate (P = .01) for the combined vectors of version and inclination. Patient-specific guides also had fewer instances of significant component malposition than standard instrumentation did. Patient-specific targeting guides were more accurate than traditional instrumentation and had fewer instances of component malposition for glenoid component placement in this multi-surgeon cadaver study of arthritic shoulders. Long-term clinical studies are needed to determine if these improvements produce improved functional outcomes.

Medical subject headings

Anatomy