Complications after arthroscopic coracoclavicular reconstruction using a single adjustable-loop-length suspensory fixation device in acute acromioclavicular joint dislocation.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 25543250.
- Also identified by DOI 10.1016/j.arthro.2014.11.013.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The purpose of this study was to evaluate clinical and radiological outcomes after arthroscopically assisted coracoclavicular (CC) fixation using a single adjustable-loop-length suspensory fixation device for acute acromioclavicular dislocation and to report intraoperative and postoperative complications. Eighteen consecutive patients with acute acromioclavicular dislocation underwent arthroscopically assisted CC fixation using a single TightRope (Arthrex, Naples, FL). Using the Rockwood classification, 3 patients had grade III dislocations, one patient had a grade IV dislocation, and 14 patients had grade V dislocations. The preoperative CC distance of the injured shoulder was 16.1 ± 2.7 mm (range, 11.2 to 21.0 mm), and it increased by 99% ± 36% (range, 17% to 153%) on average compared with the contralateral shoulder. The average CC distance was 10.5 ± 2.5 mm (range, 7.7 to 15.5 mm), and it increased by 30% ± 30% (range, -9.4% to 90%) at the final follow-up. Compared with immediate postoperative radiographs, the CC distance was maintained in 12 patients, increased between 50% and 100% in 4 patients, and increased more than 100% in 2 patients at final follow-up. However, there was no statistical difference in Constant scores between 6 patients with reduction loss (95.6 ± 4.5) and 12 patients with reduction maintenance (98.4 ± 2.5; P = .17). Perioperative complications occurred in 8 patients, including one case of acromioclavicular arthritis, one case of delayed distal clavicular fracture at the clavicular hole of the device, 3 cases of clavicular or coracoid button failures, and 3 cases of clavicular bony erosion. Satisfactory clinical outcomes were obtained after CC fixation using the single adjustable-loop-length suspensory fixation device for acute acromioclavicular joint dislocation. However, CC fixation failure of greater than 50% of the unaffected side in radiological examinations occurred in 33% of the patients within 3 months after the operation. Additionally, 8 patients (44%) had complications associated with the adjustable-loop-length suspensory fixation device and surgical technical problems. Despite acceptable shoulder function restoration, adequate care should be exercised in surgical treatment of acute acromioclavicular dislocation with a single adjustable-loop-length suspensory fixation device for optimal radiological outcomes. Level IV, therapeutic case series.
Medical subject headings
- Acromioclavicular Joint
- Clavicle
- Intraoperative Complications
- Joint Dislocations
- Orthopedic Procedures
- Postoperative Complications
- Plastic Surgery Procedures
Anatomy
- clavicle