Initial clinical experience with the Symphony heart assist system.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 25555946.
- Also identified by DOI 10.1016/j.athoracsur.2014.07.094.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Current cardiac assist devices provide full support, require a major operation, and function asynchronously to the native heart. In contrast, we developed a novel circulatory support device that provides synchronous partial support and can be placed with a minor operation. We report the first clinical implantation with the Symphony device (Abiomed, Danvers, MA). Patients with advanced heart failure despite optimal therapy who had exhausted all options were evaluated. A 64-year-old man with ischemic cardiomyopathy underwent implantation of the Symphony device in the right infraclavicular fossa. After initiating device support, the cardiac index increased from 1.7 to 2.5 L/min/m(2), pulmonary capillary wedge pressure decreased from 26 to 13 mm Hg, right atrial pressure decreased from 12 to 7 mm Hg, creatinine level decreased from 2.3 to 1.5 mg/dL, and New York Heart Association (NYHA) class improved from IIIB to II. Placement of the Symphony device resulted in improvements in hemodynamics and functional status. Further clinical data will help define the role for this approach of partial synchronous support through a less invasive operation in patients with advanced heart failure.
Medical subject headings
- Heart Failure
- Heart-Assist Devices
- Myocardial Ischemia