Validation of a computer-assisted method for measurement of radiographic wear in total hip arthroplasty using all polyethylene cemented acetabular components.

Langlois, Jean; Zaoui, Amine; Scemama, Caroline; Martell, John; Bragdon, Charles; Hamadouche, Moussa · J Orthop Res · 2015

biomechanical · Level V

Where this comes from

Abstract

Although cemented all polyethylene (PE) cups have been routinely used in total hip arthroplasty for decades, no computer-assisted method for measurement of radiographic wear has ever been specifically validated for these implants. Using a validated hip phantom model, AP plain hip radiographs were obtained consecutively for eight simulated wear positions. A version of Martell's Hip Analysis Suite software dedicated to all polyethylene sockets was used by three different examiners of varied experience. Bias (mean, standard deviation and 95% confidence interval limit), repeatability (standard deviation and 95% limit) and reproducibility (standard deviation and 95% limit) for two-dimensional wear measurements were assessed, as recommended by the current ASTM guidelines. Using this protocol, the dedicated software showed an overall mean bias of 0.089 ± 0.060 mm (mean ± SD), and 0.118 mm for 95% CI limit. Repeatability (intra examiner) standard deviation and 95% limit were respectively 0.106 mm and 0.292 mm. Reproducibility (inter examiner) standard deviation and 95% limit were respectively 0.112 mm and 0.308 mm. Martell Hip Analysis for all PE cemented cups is a reliable and low-cost instrument in the assessment of wear, despite being less precise than its original version dedicated to cementless components.