Revision of Recalled Modular Neck Rejuvenate and ABG Femoral Implants.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 25573180.
- Also identified by DOI 10.1016/j.arth.2014.12.002.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Modular neck femoral stems have been associated with adverse local tissue reactions (ALTR), leading to a voluntary recall, but these effects have not been well-characterized. A retrospective review of intraoperative findings and cobalt/chromium levels was performed in 103 hips undergoing revision for ALTR. The average preoperative serum cobalt level was 7.6 μg/L (range 1.1-23 μg/L) and chromium level was 1.8 μg/L (range 0.1-6.8 μg/L). Metallic sludge was noted in 100%, synovitis in 98%, pericapsular rind in 82%, and calcar erosion in 85%. An osteotomy was required for removal in 44%. We concluded that revision of modular neck femoral stems is associated with increased preoperative metal ion levels and stem-neck corrosion. Despite advanced stem explantation techniques, osteotomy was frequently required, leading to increased morbidity.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Femur Neck
- Hip Prosthesis
Anatomy
- femur
- hip