Evaluation of the utility of the Vigileo FloTrac(™) , LiDCO(™) , USCOM and CardioQ(™) to detect hypovolaemia in conscious volunteers: a proof of concept study.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 25583188.
- Also identified by DOI 10.1111/anae.12949.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
It is important to detect and treat hypovolaemia; however, detection is particularly challenging in the conscious, spontaneously breathing patient. Eight healthy male volunteers were monitored using four minimally invasive monitors: Vigileo FloTrac(™) ; LiDCOrapid(™) ; USCOM 1A; and CardioQ(™) oesophageal Doppler. Monitor output and clinical signs were recorded during incremental venesection of 2.5% estimated blood volume aliquots to a total of 20% blood volume removed. A statistically significant difference from baseline stroke volume was detected after 2.5% blood loss using the LiDCO (p = 0.007), 7.5% blood loss using the USCOM (p = 0.019), and 12.5% blood loss using the CardioQ (p = 0.046) and the FloTrac (p = 0.028). Receiver operator characteristic curves for predicting > 10% blood loss had areas under the curve of 0.68-0.82. The minimally invasive cardiac output devices tested can detect blood loss by a reduction in stroke volume in awake volunteers, and may have a role in guiding fluid replacement in conscious patients with suspected hypovolaemia.
Medical subject headings
- Cardiac Output
- Consciousness
- Hypovolemia
- Monitoring, Physiologic
- Stroke Volume