Looped suture properties: implications for multistranded flexor tendon repair.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 25698798.
- Also identified by DOI 10.1177/1753193413516245.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Multiple-strand repair techniques are commonly used to repair cut flexor tendons to achieve initial biomechanical strength. Looped sutures achieve multiple strands with fewer passes and less technical complexity. Their biomechanical performance in comparison with an equivalent repair using a single-stranded suture is uncertain. This study examined the mechanical properties of double-stranded loops of 3-0 and 4-0 braided polyester (Ticron) and polypropylene monofilament (Prolene). Double loops were generally less than twice the strength of a single loop. Ticron and Prolene had the same strengths, but Ticron was stiffer. The 4-0 double loops had significantly higher stiffness than 3-0 single loops. Increasing the size of sutures had a larger relative effect on strength than using a double-stranded suture. However, a double-strand loop had a larger effect on increasing stiffness than using a single suture of a larger equivalent size. Looped suture repairs should be compared with standard techniques using a thicker single suture.
Medical subject headings
- Sutures
- Tendon Injuries