Dynamic contrast-enhanced imaging of the wrist in rheumatoid arthritis: dedicated low-field (0.25-T) versus high-field (3.0-T) MRI.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 25715838.
- Also identified by DOI 10.1007/s00256-015-2120-x.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To compare the assessment of wrist synovitis severity, synovial volume and synovial perfusion parameters on a dedicated low-field (0.25-T) to that of a high-field (3-T) whole-body MR system in patients with rheumatoid arthritis (RA). Twenty-one patients (mean age 50.0 ± 9.8 years) with active RA were recruited prospectively. Dynamic contrast-enhanced MRI examination of the most severely affected wrist was performed at both 0.25 T and 3 T. Three MRI-derived parameters, synovitis severity (RAMRIS grade), synovial volume (ml(3)) and synovial perfusion indices (maximum enhancement and enhancement slope), were compared. Comparing 0.25- and 3-T MRI, there was excellent agreement for semiquantitative assessment (r: 0.80, p < 0.00001) of synovitis (RAMRIS) as well as quantitative assessment (r: 0.94, p < 0.00001) of synovial volume. Good agreement for synovial Emax (r: 0.6, p = 0.002) and fair agreement (r: 0.5, p = 0.02) for synovial Eslope was found. Imaging of the RA wrist at 0.25 T yields excellent correlation with 3 T with regard to the synovitis activity score (RAMRIS) and synovial volume measurement. Fair to good correlation between low- (0.25-T) and high-field (3-T) MR systems was found for perfusion parameters, being better for Emax than for Eslope.
Medical subject headings
- Arthritis, Rheumatoid
- Image Enhancement
- Magnetic Resonance Imaging
- Meglumine
- Organometallic Compounds
- Synovitis
- Wrist Joint
Anatomy
- wrist