Trunnion-Head Stresses in THA: Are Big Heads Trouble?
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 25724112.
- Also identified by DOI 10.1016/j.arth.2015.01.021.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The effects of large heads on stresses at the THA trunnion-head junction and their impact on tribocorrosion/metal ion release remain controversial. A 12/14 3D-model of a stem with different head sizes was investigated. Material properties of titanium were assigned to the trunnion and cobalt-chrome/alumina to the heads. A load simulating walking single-leg stand phase was applied to the head. A total contact head-trunnion interface was assumed. The area underneath the junction underwent significant elevations in stresses as head size increased from 28- to 40-mm. Maximum principal stress doubled between 28 and 40-mm heads, regardless of head material. Stress levels had a direct correlation to head diameter. Stress increases observed using increasingly larger heads will probably contribute to head-trunnion tribocorrosion and ion release.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
Anatomy
- femur
- hip