Effect of Taper Design on Trunnionosis in Metal on Polyethylene Total Hip Arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 25773576.
- Also identified by DOI 10.1016/j.arth.2015.02.031.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study examines how taper design affects corrosion and fretting at the head trunnion surface. All hip prostheses retrieved between 1999 and 2013 with 28mm/+0 heads were selected, resulting in 44 cobalt-chrome-on-polyethylene implants, representing six taper designs. Mean implantation time: 8.9±3.7years. The femoral head tapers were scored for fretting and corrosion using the Goldberg scale as both a combined score and by three zones (apex, central and base). There was no difference in age (P=0.34), BMI (P=0.29), or implantation time (P=0.19) between taper groups. The 11/13 taper had the highest combined corrosion and fretting score, but no difference (P=0.22) between groups for combined scores (P=0.22 for corrosion, P=0.19 for fretting). In a zone-specific analysis, the 11/13 taper had highest corrosion score at base zone (P=0.02). Taper design had a significant effect on corrosion at base of trunnion.