Field Reliability Estimation for Cochlear Implants.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 25775485.
- Also identified by DOI 10.1109/TBME.2015.2412127.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Cochlear implants are electronic devices used to provide useful hearing sensations to patients with severe to profound hearing loss. Safety and reliability are considered by patients and clinicians when selecting among the different models of cochlear implants. However, comparing reliability estimates from different manufacturers is difficult because of the lack of a standardized approach for classifying and quantifying failure data. This paper develops a classification scheme for cochlear implant failures wherein each failure report is categorized based on the failure site, cause, mode, and mechanism obtained through failure modes, mechanisms, and effects analysis. Failure parameters based on the Weibull distribution are used to quantify the following reliability metrics: B1, B5, B10, and B50 lives. The developed reporting scheme provides clear and quantitative information on failure causes and failure sites for a given implant. This method provides a consistent reporting format for communicating field reliability to regulatory units, clinicians, and patients.
Medical subject headings
- Cochlear Implants
- Equipment Failure Analysis