Radiofrequency ablation near the bone-muscle interface alters soft tissue lesion dimensions.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 25785840.
- Also identified by DOI 10.1097/AAP.0000000000000221.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Radiofrequency (RF) lesions are safe and effective in the treatment of spine pain; however, models developed to study factors affecting lesion dimensions have been performed in homogeneous media that may not accurately simulate human anatomy and electrophysiology. We present a novel ex vivo porcine model for performing RF lesion studies and report the influence of bone on projection of RF ablation lesions into soft tissue. Radiofrequency lesions were performed in porcine rib specimens using monopolar 18-gauge, 10-mm straight active tip cannula, with a lesion temperature of 80°C for 150 seconds. Ten lesions were performed in pure porcine muscle tissue and abutting porcine rib bone with surrounding muscle. Lesions were exposed with dissection and measured with digital calipers. Maximal effective lesion radius approximately doubled against the bone compared with the pure muscle group (mean, 5.65 mm [95% CI, 5.43-5.87 mm] vs 2.68 mm [95% CI, 2.55-2.81 mm], P < .0001), although this was seen only in a vertical direction and not horizontally. In addition, the prelesion and postlesion impedance of the bone-muscle interface was consistently higher than the muscle-only interface (mean, 165.6 Ohm [95% CI, 146.6-184.6 Ohm] vs 137.8 Ohm [95% CI, 135.5-140.1 Ohm], P = 0.004; 144.3 Ohm [95% CI, 134.3-154.3 Ohm] vs 124.3 Ohm [95% CI, 119.3-129.3 Ohm], P = 0.001). Other dimensions and estimated volume were not significantly different. Bone adjacent to RF lesions alters the surrounding electrophysiological environment causing RF lesions to project further perpendicularly from the needle axis, vertically to bone, than previously expected. This phenomenon should be considered in the future modeling and clinical practice of RF.
Medical subject headings
- Body Weights and Measures
- Catheter Ablation
- Hot Temperature
- Muscle, Skeletal