Usefulness of metal artifact reduction with WARP technique at 1.5 and 3T MRI in imaging metal-on-metal hip resurfacings.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 25805436.
- Also identified by DOI 10.1007/s00256-015-2128-2.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To evaluate the usefulness of the metal artifact reduction technique "WARP" in the assessment of metal-on-metal hip resurfacings at 1.5 and 3T in the context of image quality and imaging speed. Nineteen patients (25 hip resurfacings) were randomized for 1.5 and 3T MRI, both including T1 and T2 turbo spin-echo as well as turbo inversion recovery magnitude sequences with and without view angle tilting and high bandwidth. Additional 3T sequences were acquired with a reduced number of averages and using the parallel acquisition technique for accelerating imaging speed. Artifact size (diameter, area), image quality (5-point scale) and delineation of anatomical structures were compared among the techniques, sequences and field strengths using the Wilcoxon sign-rank and paired t-test with Bonferroni correction. At both field strengths, WARP showed significant superiority over standard sequences regarding image quality, artifact size and delineation of anatomical structures. At 3T, artifacts were larger compared to 1.5T without affecting diagnostic quality, and scanning time could be reduced by up to 64 % without quality degradation. WARP proved useful in imaging metal-on-metal hip resurfacings at 1.5T as well as 3T with better image quality surrounding the implants. At 3T imaging could be considerably accelerated without losing diagnostic quality.
Medical subject headings
- Artifacts
- Hip Joint
- Hip Prosthesis
- Image Enhancement
- Image Interpretation, Computer-Assisted
- Magnetic Resonance Imaging
- Metal-on-Metal Joint Prostheses
- Metals
Anatomy
- hip