Micro-scale testing and micromechanical modelling for high cycle fatigue of CoCr stent material.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 25817609.
- Also identified by DOI 10.1016/j.jmbbm.2015.02.011.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This paper presents a framework of experimental testing and crystal plasticity micromechanics for high cycle fatigue (HCF) of micro-scale L605 CoCr stent material. Micro-scale specimens, representative of stent struts, are manufactured via laser micro-machining and electro-polishing from biomedical grade CoCr alloy foil. Crystal plasticity models of the micro-specimens are developed using a length scale-dependent, strain-gradient constitutive model and a phenomenological (power-law) constitutive model, calibrated from monotonic and cyclic plasticity test data. Experimental microstructural characterisation of the grain morphology and precipitate distributions is used as input for the polycrystalline finite element (FE) morphologies. Two microstructure-sensitive fatigue indicator parameters are applied, using local and non-local (grain-averaged) implementations, for the phenomenological and length scale-dependent models, respectively, to predict fatigue crack initiation (FCI) in the HCF experiments.
Medical subject headings
- Alloys
- Chromium
- Cobalt
- Finite Element Analysis
- Materials Testing
- Mechanical Phenomena
- Stents