An Implantable RFID Sensor Tag toward Continuous Glucose Monitoring.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 25823049.
- Also identified by DOI 10.1109/JBHI.2015.2415836.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This paper presents a wirelessly powered implantable electrochemical sensor tag for continuous blood glucose monitoring. The system is remotely powered by a 13.56-MHz inductive link and utilizes an ISO 15693 radio frequency identification (RFID) standard for communication. This paper provides reliable and accurate measurement for changing glucose level. The sensor tag employs a long-term glucose sensor, a winding ferrite antenna, an RFID front-end, a potentiostat, a 10-bit sigma-delta analog to digital converter, an on-chip temperature sensor, and a digital baseband for protocol processing and control. A high-frequency external reader is used to power, command, and configure the sensor tag. The only off-chip support circuitry required is a tuned antenna and a glucose microsensor. The integrated chip fabricated in SMIC 0.13-μm CMOS process occupies an area of 1.2 mm ×2 mm and consumes 50 μW. The power sensitivity of the whole system is -4 dBm. The sensor tag achieves a measured glucose range of 0-30 mM with a sensitivity of 0.75 nA/mM.
Medical subject headings
- Blood Glucose
- Monitoring, Physiologic
- Prostheses and Implants
- Radio Frequency Identification Device
- Wireless Technology