Sensitivity and specificity of the AdenoPlus point-of-care system in detecting adenovirus in conjunctivitis patients at an ophthalmic emergency department: a diagnostic accuracy study.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 25824258.
- Also identified by DOI 10.1136/bjophthalmol-2014-306508.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To estimate the diagnostic accuracy (sensitivity and specificity) of the AdenoPlus point-of-care adenoviral test compared to PCR in an ophthalmic accident and emergency service. These findings were compared with those of a previous study. This was a prospective diagnostic accuracy study on 121 patients presenting to an emergency eye unit with a clinical picture of acute adenoviral conjunctivitis. AdenoPlus testing was carried out on one eye of each patient and a PCR analysis was also performed on a swab taken from the same eye. AdenoPlus and PCR results were interpreted by masked personnel. Sensitivity and specificity for the AdenoPlus test were calculated using PCR results as the reference standard. 121 patients were enrolled and 109 met the inclusion criteria. 43 patients (39.4%) tested positive for adenovirus by PCR analysis. The sensitivity of the AdenoPlus swab in detecting adenovirus was 39.5% (17/43, 95% CI 26% to 54%) and specificity was 95.5% (63/66, 95% CI 87% to 98%) compared to PCR. The AdenoPlus test has a high specificity for diagnosing adenoviral conjunctivitis, but in this clinical setting, we could not reproduce the high sensitivity that has been previously published.
Medical subject headings
- Adenovirus Infections, Human
- Conjunctivitis, Viral
- Diagnostic Techniques, Ophthalmological
- Emergency Service, Hospital
- Point-of-Care Systems