Arrhythmia discrimination using a smart phone.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 25838530.
- Also identified by DOI 10.1109/JBHI.2015.2418195 and PMC identifier 6599713.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We hypothesize that our smartphone-based arrhythmia discrimination algorithm with data acquisition approach reliably differentiates between normal sinus rhythm (NSR), atrial fibrillation (AF), premature ventricular contractions (PVCs) and premature atrial contraction (PACs) in a diverse group of patients having these common arrhythmias. We combine root mean square of successive RR differences and Shannon entropy with Poincare plot (or turning point ratio method) and pulse rise and fall times to increase the sensitivity of AF discrimination and add new capabilities of PVC and PAC identification. To investigate the capability of the smartphone-based algorithm for arrhythmia discrimination, 99 subjects, including 88 study participants with AF at baseline and in NSR after electrical cardioversion, as well as seven participants with PACs and four with PVCs were recruited. Using a smartphone, we collected 2-min pulsatile time series from each recruited subject. This clinical application results show that the proposed method detects NSR with specificity of 0.9886, and discriminates PVCs and PACs from AF with sensitivities of 0.9684 and 0.9783, respectively.
Medical subject headings
- Arrhythmias, Cardiac
- Cell Phone
- Mobile Applications
- Signal Processing, Computer-Assisted