A new technique to explant an infected aortic endograft.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 25937607.
- Also identified by DOI 10.1016/j.jvs.2015.03.033.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The management of an infected aortic endograft can be challenging both operatively and clinically. Although aortic endograft infection is rare, the incidence is likely to increase in the coming years because of ever rising numbers of endovascular aneurysm repairs. Definitive management involves the removal of the endograft through laparotomy. Removal of the graft is technically challenging; no manufacturer's device is available to assist in disengagement of barbed hooks that hold the endograft in position. We present a new technique using the disposable proctoscope as a device to facilitate safe removal of the endograft with minimal damage to the aortic wall.
Medical subject headings
- Aortic Aneurysm, Abdominal
- Blood Vessel Prosthesis
- Blood Vessel Prosthesis Implantation
- Device Removal
- Prosthesis-Related Infections
- Stents