MRI Based Comparison of Tibial Bone Coverage by Five Knee Prosthesis: Anthropometric Study in Indians.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 25956525.
- Also identified by DOI 10.1016/j.arth.2015.03.041.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This MRI based study evaluates morphological differences of proximal tibia (total cross-sectional area, mediolateral and anteroposterior distance) 8-10 mm distal to the lateral tibial plateau. We evaluated the difference in the coverage of the tibial surface between symmetric and asymmetric tibial trays and difference in coverage between males and females. 150 patients who underwent MRI scans for sports related soft tissue injury without osteoarthritis were studied. The tibial trays of the 5 total knee arthroplasty designs (4 symmetric and 1 asymmetric) were scanned. Mean total tibial coverage of all designs was more than 80%. Asymmetric baseplate had maximum total tibial coverage and maximum rate of optimal fit, with only 2% absolute overhang posterolaterally. Females had better tibial coverage as compared to males.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Knee Joint
- Magnetic Resonance Imaging
- Tibia
Anatomy
- knee
- tibia