Determination of gas flow through airway exchange catheters: measured and calculated values and dependence on pressure and entrainment.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 25959305.
- Also identified by DOI 10.1111/anae.13109.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Serious complications have been described during oxygenation of patients with airway exchange catheters, due to catheter malpositioning, accidentally applied high airway pressures or high delivered volumes. In this in-vitro study, we analysed gas flow through various airway exchange catheters and described its dependence on driving pressure and entrainment. We applied driving pressures between 0.5 and 2.5 bar and observed maximal flow rates of 3.6 l.s(-1) . Measured gas flow values differed significantly from values calculated according to the Hagen-Poiseuille equation. Although flow restriction in ventilators and small-bore connectors will limit gas flow, large gas volumes may be unintentionally applied via the airway exchange catheters, leading to serious complications.
Medical subject headings
- Catheters
- Respiration, Artificial
- Ventilators, Mechanical