Comparison of cannulated screws with FiberWire or stainless steel wire for patella fracture fixation: A pilot study.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 25972700.
- Also identified by DOI 10.1016/j.jor.2014.04.011 and PMC identifier 4421085.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This pilot study tested whether FiberWire provides similar protection to steel wire against repair displacement in patella fractures. Thirteen cadaver knees were cyclically loaded with 10 cycles (0-90° flexion) and fracture displacement was recorded. Fixation methods were also tested in load to failure (>3 mm displacement). There was no difference between wire types in fracture displacement (1.4 mm ± 0.33 mm vs 1.2 mm ± 0.34 mm, respectively; p = 0.418) or in the load to failure (714.7 N ± 110.9 N vs 744.5 N ± 92.8 N, respectively; p = 0.360). FiberWire provides similar protection to steel wire against repair displacement after fixation of patellar fractures.
Anatomy