Comparison of Acetabular Bone Resection, Offset, Leg Length and Post Operative Function Between Hip Resurfacing Arthroplasty and Total Hip Arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 25981328.
- Also identified by DOI 10.1016/j.arth.2015.04.030.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Controversy exists regarding the amount of acetabular bone resection, biomechanics and function of patients receiving either total hip arthroplasty (THA) or hip resurfacing arthroplasty (HRA). A cohort of patients undergoing 36 mm ceramic-on-ceramic THA (89) or metal-on-metal HRA (86) were compared. No difference was observed when the ratio of native femoral head size was compared to the implanted acetabular component size (1.15 ± 0.1 HRA c.f. 1.13 ± 0.1 THA). No difference was observed in acetabular offset, vertical centre of rotation or function (OHS mean 47 in both groups) but leg length discrepancy (1.8 mm c.f. 5.5 mm) and femoral offset did differ (0.6 mm c.f. 4.1 mm). This demonstrates that 36 mm ceramic-on-ceramic THA is not associated with more bone resection than HRA and achieves equivalent function whilst avoiding the problems of metal-on-metal bearings.
Medical subject headings
- Acetabulum
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Leg
- Leg Length Inequality