Poor accuracy of plain radiographic measurements of prosthetic migration and alignment in total ankle replacement.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 25986485.
- Also identified by DOI 10.1186/s13018-015-0220-x and PMC identifier 4490685.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
The rotational position of the leg substantially influences measurements in ankle radiographs after total ankle replacement (TAR). The aim of our study was to further specify the influence of different projections on radiographic parameters used to assess component position after TAR. The effect of varying degrees of internal rotation, flexion, and ab-/adduction on reference lines in anteroposterior and lateral ankle radiographs was investigated in a cadaveric TAR model. Observed variations were then compared with those found in 34 consecutive patients that received a HINTEGRA total ankle prosthesis in our department. A change of rotation of 20° resulted in a variation of measured reference lines of more than 1.3 cm in anteroposterior radiographs and more than 1 cm in lateral radiographs in our experimental setting. Even higher intraindividual changes of up to 1.4 cm were observed in our in vivo series. The findings suggest that rotational position of the leg highly influences measurements in ankle radiographs after TAR. It further raises the question, if previously described radiographic parameters do provide accurate information for the outcome after TAR in clinical routine as suggested in literature.
Medical subject headings
- Ankle Joint
- Arthroplasty, Replacement, Ankle
- Prosthesis Failure
Anatomy
- ankle