Effectiveness of a load-imposing device for cyclic stretching of isolated human bronchi: a validation study.
other · Level V
Where this comes from
- Record sourced from PubMed, PMID 26011598.
- Also identified by DOI 10.1371/journal.pone.0127765 and PMC identifier 4444237.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
BACKGROUND: Mechanical ventilation may induce harmful effects in the airways of critically ill patients. Nevertheless, the effects of cyclic stretching caused by repetitive inflation-deflation of the bronchial compartment have not been well characterized in humans. The objective of the present study was to assess the effectiveness of a load-imposing device for the cyclic stretching of human bronchi. METHODS: Intact bronchial segments were removed from 128 thoracic surgery patients. After preparation and equilibration in an organ bath, bronchi were stretched repetitively and cyclically with a motorized transducer. The peak force imposed on the bronchi was set to 80% of each individual maximum contraction in response to acetylcholine and the minimal force corresponded to the initial basal tone before stretching. A 1-min cycle (stretching for 15 sec, relaxing for 15 sec and resting for 30 sec) was applied over a time period ranging from 5 to 60 min. The device's performance level was assessed and the properties of the stretched bronchi were compared with those of paired, non-stretched bronchi. RESULTS: Despite the intrinsic capacities of the device, the targets of the tension adjustments remained variable for minimal tension (156-178%) while the peak force set point was unchanged (87-115%). In the stretched bronchi, a time-dependent rise in basal tone (P < .05 vs. non-stretched) was apparent after as little as 5 min of cyclic stretching. The stretch-induced rise in basal tone continued to increase (P < .01) after the stretching had ended. Only 60 min of cyclic stretching was associated with a significant (P < .05) increase in responsiveness to acetylcholine, relative to non-stretched bronchi. CONCLUSIONS: Low-frequency, low-force, cyclic loading of human bronchi is associated with elevated basal tone and acetylcholine responsiveness. The present experimental model is likely to be a useful tool for future investigations of the bronchial response to repetitive stress during mechanical ventilation.
Medical subject headings
- Bronchi
- Tissue Expansion Devices