Laryngoscope Illuminance in a Tertiary Care Medical Center: Industry Standards and Implications for Quality Laryngoscopy.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 26092842.
- Also identified by DOI 10.1177/0194599815587700.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To test the hypothesis that a substantial proportion of laryngoscopes exhibit substandard illuminance by comparing laryngoscope illuminance in a tertiary-level medical center to established standards and identifying features associated with poor illuminance. Cross-sectional observational study. Academic tertiary care medical center (level 1 trauma center, specialty cardiac hospital, and general hospital). Laryngoscopes from main, cardiac, and outpatient operating rooms; emergency department; and code carts were tested using a standard technique. Illuminance (lux) was chosen as the outcome measure. Benchmarks were derived from the International Standards Organization and medical literature. Light types included incandescent bulb, light-emitting diode, and xenon. Personnel were surveyed regarding maintenance practices. Across all hospitals, 691 laryngoscopes were tested. Mean (SD) illuminance was 810 (700) lux for incandescent bulb-on-blade designs (n = 237), 1860 (1220) lux for incandescent bulb in-handle designs (n = 79), 4730 (3210) lux for LED (n = 354), and 28,800 (34,500) lux for xenon (n = 21). Seven percent of units failed to turn on (n = 45). Using an established threshold of 867 lux, 28% of devices (47% of incandescent, 12% of LED, and 10% of xenon) were substandard. All laryngoscopes were cleaned according to standard protocols following use; no preventive maintenance was reported. Twenty-eight percent of laryngoscopes in a tertiary care hospital exhibit substandard illuminance; these results corroborate the findings of our inaugural study on this subject. Consequently, our hospital is instituting changes to reduce the likelihood of substandard performance by laryngoscopes in circulation.
Medical subject headings
- Industry
- Laryngoscopes
- Laryngoscopy
- Lighting
- Operating Rooms
- Tertiary Care Centers