The lubrication performance of the ceramic-on-ceramic hip implant under starved conditions.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 26114217.
- Also identified by DOI 10.1016/j.jmbbm.2015.06.001 and PMC identifier 4570925.
- Licence recorded as CC BY.
- The licence permits redistribution, so the abstract is shown in full and the full text is available from the publisher.
Abstract
Lubrication plays an important role in the clinical performance of the ceramic-on-ceramic (CoC) hip implant in terms of reducing wear and avoiding squeaking. All the previous lubrication analyses of CoC hip implants assumed that synovial fluid was sufficiently supplied to the contact area. The aim of this study was to investigate the lubrication performance of the CoC hip implant under starved conditions. A starved lubrication model was presented for the CoC hip implant. The model was solved using multi-grid techniques. Results showed that the fluid film thickness of the CoC hip implant was affected by fluid supply conditions: with the increase in the supplied fluid layer, the lubrication film thickness approached to that of the fully blooded solution; when the available fluid layer reduced to some level, the fluid film thickness considerably decreased with the supplying condition. The above finding provides new insights into the lubrication performance of hip implants.
Medical subject headings
- Ceramics
- Hip Prosthesis
- Lubrication