Minimum 10-Year Wear Analysis of Highly Cross-Linked Polyethylene in Cementless Total Hip Arthroplasty.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 26231076.
- Also identified by DOI 10.1016/j.arth.2015.06.053.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Fifty-four patients (64 hips) underwent cementless total hip arthroplasty between 2000 and 2003 with a 22-mm zirconia ceramic bearing on highly cross-linked polyethylene, and were evaluated with a mean 11.9-year postoperative follow-up (range, 10-14 years). Linear wear was measured on the anteroposterior radiograph of the hip. No evidence of osteolysis and loosening was found on the final radiograph in any of the cases, and the steady-state linear wear rate was 0.017±0.018 mm/year. No significant correlation was found between the linear wear rate and age, body weight, cup inclination angle, or polyethylene thickness. Highly cross-linked polyethylene showed excellent wear resistance for >10 years when used in combination with 22-mm zirconia heads.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
- Polyethylene
Anatomy
- hip