Biomechanical testing of circumferential instrumentation after cervical multilevel corpectomy.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 26233243.
- Also identified by DOI 10.1007/s00586-015-4167-8.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Biomechanical investigation. This study describes ex vivo evaluation of the range of motion (ROM) to characterize the stability and need for additional dorsal fixation after cervical single-level, two-level or multilevel corpectomy (CE) to elucidate biomechanical differences between anterior-only and supplemental dorsal instrumentation. Twelve human cervical cadaveric spines were loaded in a spine tester with pure moments of 1.5 Nm in lateral bending (LB), flexion/extension (FE), and axial rotation (AR), followed by two cyclic loading periods for three-level corpectomies. After each cyclic loading session, flexibility tests were performed for anterior-only instrumentation (group_1, six specimens) and circumferential instrumentation (group_2, six specimens). The flexibility tests for all circumferential instrumentations showed a significant decrease in ROM in comparison with the intact state and anterior-only instrumentations. In comparison with the intact state, supplemental dorsal instrumentation after three-level CE reduced the ROM to 12% (±10%), 9% (±12%), and 22% (±18%) in LB, FE, and AR, respectively. The anterior-only construct outperformed the intact state only in FE, with a significant ROM reduction to 57% (±35 %), 60% (±27%), and 62% (±35%) for one-, two- and three-level CE, respectively. The supplemental dorsal instrumentation provided significantly more stability than the anterior-only instrumentation regardless of the number of levels resected and the direction of motion. After cyclic loading, the absolute differences in stability between the two instrumentations remained significant while both instrumentations showed a comparable increase of ROM after cyclic loading. The large difference in the absolute ROM of anterior-only compared to circumferential instrumentations supports a dorsal support in case of three-level approaches.
Medical subject headings
- Cervical Vertebrae
- Internal Fixators
- Spinal Cord Compression
Anatomy
- cervical spine