Late Cardiac Tamponade as a Result of Parietal Pericardium Erosion.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 26234848.
- Also identified by DOI 10.1016/j.athoracsur.2014.09.085.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Late onset cardiac tamponade is a rare and particularly challenging (both from diagnostic and management perspectives) complication of intracardiac lead implantation. We present a case of a late tamponade leading to cardiogenic shock, which occurred 1,164 days after implantable cardioverter-defibrillator (ICD) implantation. Open repair revealed unusual and, to our knowledge, not yet reported mechanism of the disease. A pressure sore caused by an ICD lead was found in the parietal layer of pericardium with no visible damage to the visceral layer. Conservative management in the described clinical scenario could be fatal, thus awareness of this pathomechanism of tamponade is critical.
Medical subject headings
- Cardiac Tamponade
- Defibrillators, Implantable
- Pericardium
- Postoperative Complications