Comparison of Tibial Insert Polyethylene Damage in Rotating Hinge and Highly Constrained Total Knee Arthroplasty: A Retrieval Analysis.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 26253478.
- Also identified by DOI 10.1016/j.arth.2015.07.007.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
This study compared the damage scores and damage patterns in 19 tibial inserts from rotating hinge (RH) implants with 19 inserts from highly constrained (HC) implants. Each insert was divided into 16 damage zones and each zone was subjectively graded from a scale of 0-3 for seven different damage modes. The overall damage scores were comparable for the two groups (RH: 64.1 ± 15.4; HC: 66.1 ± 29.0; P = 0.59). The HC group, however, had greater post damage (compared to the post-hole of RH) while the RH group had greater backside damage. The pattern of damage was also different, with burnishing and cold flow being more common in HC group while pitting, scratching and embedded debris were more common in the RH group.
Medical subject headings
- Arthroplasty, Replacement, Knee
- Knee Prosthesis
- Polyethylene
- Prosthesis Design
- Tibia
Anatomy
- knee
- tibia