Stability of Uncemented Cups - Long-Term Effect of Screws, Pegs and HA Coating: A 14-Year RSA Follow-Up of Total Hip Arthroplasty.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 26260783.
- Also identified by DOI 10.1016/j.arth.2015.07.012.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Screws, pegs and hydroxyapatite-coating are used to enhance the primary stability of uncemented cups. We present a 14-year follow-up of 48 hips randomized to four groups: press-fit only, press-fit plus screws, press-fit plus pegs and hydroxyapatite-coated cups. Radiostereometric migration measurements showed equally good stability regardless cup augmentation. The mean wear rate was high, 0.21 mm/year, with no differences between the groups. Seven hips had radiographical osteolysis but only in hips with augmented cups. Cups without screw-holes compared with cups with screw-holes resulted in better clinical outcome at the 14-year follow-up. Thus, augmentation of uncemented cups with screws, pegs, or hydroxyapatite did not appear to improve the long-term stability compared with press-fit only.
Medical subject headings
- Arthroplasty, Replacement, Hip
- Hip Prosthesis
Anatomy
- hip