Biomechanical Assessment of Locked Plating for the Fixation of Patella Fractures.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 26299810.
- Also identified by DOI 10.1097/BOT.0000000000000309.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To analyze the mechanical stability of locked plating in comparison with tension-band wiring for the fixation of fractures of the patella. Biomechanical tests were performed on artificial foam patella specimens comparing an angular stable plate and monocortical screws with tension-band wiring. Tests were performed under combined tension and bending until failure simulating physiological loading of the tibia during walking. Tension-band wiring failed at 66% of the failure load of plating (1052 N, P = 0.002) and had 5 times larger fracture gap displacements (P = 0.002). Based on the biomechanical advantages, locked plating of the patella may constitute a reasonable alternative in the treatment of patella fractures.
Medical subject headings
- Bone Plates
- Fracture Fixation, Internal
- Fractures, Bone
- Patella
Anatomy
- knee